5
2.2 Transport and storage conditions
1Storage temperature: -40~55.
2Relative humidity: ≤93%.
3. Installation
This section describes the instructions that you should abide when install the centrifuge to ensure your
safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
WARNING
Improper power supply may damage centrifuge.
Make sure the power source conforms to the required power supply before connecting.
3.1 Location
1Place this centrifuge on a firmflat and level surface, ensure the four feet of this centrifuge stand on
the counter firmly. Avoid installing on a slippery surface or surface prone to vibration.
2Ideal ambient temperature is 20±5 , avoid placing the centrifuge in direct sunlight if temperature
exceeds 30.
3Keep clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to guarantee the
cooling efficiency.
4Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
3.2 Connection of the power cord and grounding
WARNING
To avoid electrical shocks, ensure your hands are dry when touching the power cord.
This centrifuge must be grounded properly.
An minimum 10A outlet providing a sufficient ground is required, and this must meet local safety
requirements.