6141.22X Incubator 8000 IC/SC/NC
Function Description
09.99 Page 15 of 49
i. Evaluation of valve monitoring
j. Triggering of INOP-LED
k. Triggering of additional fan
2.1.6 Climate sensor housing (Inc. 8000 IC)
This sensor detects the climatic conditions inside the incubator. For calibration of the 02
measurement function the sensor can be swung out of the incubator interior.
Function blocks:
- 2 air temperature sensors
- 2 02 sensors
- Terminal for skin temperature sensor (units with software version 20.n only)
- Humidity sensor with evaluation electronics
-
2 micro switches to determine sensor position
2.1.7 Air temperature sensor (Inc. 8000 SC / NC)
For measurement of the air temperature two identical sensors are enclosed in one housing.
2.1.8 Height adjustment
The height is adjusted via an Electro-motor with spindle. The pedestal is triggered via a foot-
operated switch. The Electro-motor is equipped with an automatically resettable over-
temperature switch.
2.2. Description of block diagram of electronic functioning
2.2.1 Incubator 8000 IC / SC / NC block diagrams