6141.22X Incubator 8000 IC/SC/NC
Function Description
09.99 Page 47 of 49
3.4 Climate sensor housing Inc. 8000 IC
3.4.1 Intended use
The Incubator sensor serves measurement of the climatic conditions inside the
Incubator 8000 IC. The air temperature, 02-concentration and the relative air humidity are
monitored. On units with software version 20.n (without ThermoMonitoring) a skin temperature
sensor may be connected to devices which are equipped accordingly.
3.4.2 Arrangement
3.4.3 Detailed description
3.4.3.1 Temperature sensors
The air temperature sensor comprises 2 NTCs.
Temperature in °C Resistance in Ohm
20 6246.0
21 5970.8
22 5709.4
23 5460.9