6141.22X Incubator 8000 IC/SC/NC
Function Description
09.99 Page 7 of 49
1.3.4 Total shutdown
Faults in the central microprocessor control system and in the air temperature control module
result in complete shutdown of the incubator. The following alarm signals are then given:
- Red "inop" LED is lit.
- Continuous alarm tone (non-suppressible).
Individual checks should be carried out to determine whether the fault appeared only
temporarily. If the incubator shows the "Inop" signal after it is switched on again, it is
inoperative. It must not be used and must be clearly marked faulty.
2 Functional description of Incubator 8000 IC / SC / NC
2.1 Description of component block diagram
- Component block diagram 8000 IC
- Component block diagram 8000 SC / NC
- Fresh air intake and 02-supply
- Unit with air heating, fan and humidifier
- Electronics module
- Climate sensor (Inc. 8000 IC only)
- Air temperature sensor (Inc. 8000 SC / NC)
-
Height adjustment
2.1.1 Component block diagram 8000 IC