Operation
114
Operating Instructions Caleo, Software 2.n
Integrated Infant Scale
During weighing, the user is guided through the sequence of
operating steps by the following prompts.
z Lift the baby off the bed.
The system waits until the scale has stabilized and has been at
rest for 3 seconds.
– 1 beep is sounded.
z Place patient back on the bed.
Weighing then proceeds.
– 1 beep is sounded.
The weighing procedure is now complete.
– The current and last weighing results are displayed on the
screen. If the measuring range has been exceeded, the
symbol is displayed.
After 10 seconds, the display reverts to the standard screen.
or
z Press rotary knob to display the standard screen.
The screen will display the last weight as advisory text for the
next 10 minutes.