Operation
115
Operating Instructions Caleo, Software 2.n
Integrated Infant Scale
Canceling the weighing procedure
Weighing is aborted if a minimum patient weight (250 g) is not
removed or added back within 60 seconds.
3 short beeps are emitted.
Weighing without tare
If the last tare weight was obtained no longer than 60 minutes
earlier and no objects have been placed on or been removed
from the bed, a new tare weight does not have to be obtained
again.
Preparing for weighing
z Set Caleo to the horizontal position. Check with the aid of
the built-in water levels – see page 48.
z Set both knobs to the horizontal position ,
see page 45.
Starting the weighing procedure
1 Activate the weighing process = press soft key.
Caleo proposes to reweigh directly without obtaining a new
tare weight.
2 Press »reweigh« key to weigh without obtaining a new tare
weight.
Place infant back on the bed.
Weighing then proceeds.
1 beep is emitted.
206
D
2
1