EasyManua.ls Logo

Dräger caleo - Page 88

Dräger caleo
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
88
Operating Instructions Caleo, Software 2.n
Using O2 Control
The display returns to the standard screen. The measured
values are displayed.
The setpoint and »set: >40 « are displayed alternately.
The yellow bar LED lights up.
If the yellow bar LED starts flashing, the setpoint has been
increased by more than 3 % or 5 % (see page 88), or other
alarms are active.
Alarms
Alarm limits can be changed in the configuration
(see page 108).
Example: if the deviation between the set and measured
O
2 concentration exceeds ±5 %*:
The screen displays the warning message
»Oxygen deviation above 5 %«,
An alarm tone sequence (5 beeps) starts,
The central alarm indicator lights up**,
1 The measured value starts flashing,
2 The red bar LED starts blinking.
The intermittent audible alarm can be silenced for 2 minutes:
3 Press key to silence intermittent audible alarm,
or
4 Press rotary knob.
The warning message remains on screen,
The intermittent alarm tone is silenced,
The central alarm indicator goes out,
1 The measured value continues to flash,
2 The red bar LED continues to blink.
When the measured value has returned to a value within the
alarm threshold of ±5 Vol.%:
The warning message disappears.
The intermittent alarm tone is cancelled.
The central alarm indicator goes out.
1 The measured value remains on screen without flashing.
2 The red bar LED goes out.
For other alarms, see "Troubleshooting – Error Messages",
page 142.
See "Alarm Description", page 161.
*The numerical values in this description are examples.
See "Configuring alarm settings", page 107.
** The central alarm indicator light may be switched off.
See "Configuring system settings", page 106.
316
234
D
3
4
1
2

Other manuals for Dräger caleo

Related product manuals