Option 1 - Saliva Sample
1. Place collector in the mouth, move briskly from side to side
throughout the complete oral cavity (cheek pouch, under tongue,
palatal area) for one minute.
2. Sampling is complete when a stain is seen on the collector.
Maximum time for sampling is 4 minutes.
Option 2 - Surface Examination
1. Remove swab from the container of the Dräger SSK 5000 and
swipe the area to be inspected for 30 seconds.
2. Return swab to container, screw closed and shake for
30 seconds.
3. Open container and remove swab.
4. Place saliva collector from the DrugTest cassette into the container
opening.
5. Carefully invert and wait for a blue stain to form.
Option 3 - Solid Matter Examination
1. Remove swab from the container of the Dräger SSK 5000 and take
a sample by quick strokes to either the inside and/or outside of the
packaging.
NOTICE
Do not immerse swab directly into the substance!
2. Return swab to container, screw closed and shake for
30 seconds.
3. Open container and remove swab.
4. Place saliva collector from the DrugTest cassette into the container
opening.
5. Carefully invert and wait for a blue stain to form.
Analysis for all Options
1. Turn on analyser with the OK key.
Display: Await "Ready to Analyse".
2. Open door.
3. Insert DrugTest cassette and cartridge.
4. Close door: Analysis will begin automatically.
5. After conclusion of the analysis, results are indicated in the display.
6. Remove DrugTest cassette and close door to prepare for next
measurement.
i
i
Dräger DrugTest
®
5000 - quick reference guide
NOTICE
In order to operate the Dräger DrugTest 5000 test system, be sure you fully understand and strictly
observe the instructions of the analyser and the test kits utilised.