Reprocessing list
Applicable to non-infectious patients.
The reprocessing list contains guiding values only.
The instructions of the hospital’s infection control
officer shall prevail and must be observed by the
user!
Before using on patients again
1 Gather the necessary device components, see
chapter on assembly and preparation.
2 Mount the parts in reverse order from
disassembly.
3 Check readiness for operation, chapter Getting
Started.
Checklist
Components which
can be reprocessed
Recommended
reprocessing
intervals
Manual
precleaning
Machine
cleaning and
thermal
disinfection
Manual disinfection Hot steam
sterilization
Surface
disinfection
Disinfection
by immersion
bath
Rigid arm for
breathing bag
Weekly Yes Yes Yes Yes Yes
Prerequisites
Locating leakages in the breathing circuit
The check must be performed once without
vaporizer and once with vaporizer. Control dial is in
the zero position.
The D-Vapor
1)
must be switched on for the leakage
test. After the leakage test has ended, switch off the
D-Vapor again.
Change to Standby mode and press the
soft key for the leakage test.
Follow the instructions on the screen.
If the system has leakage (the pressure drops):
z Check all plug connections and screw
connections for a tight seat.
z Replace missing or damaged seals. If
necessary, contact DrägerService or the
local authorized service partner.