Manual cleaning
and disinfection
Machine cleaning
and disinfection
Sterilization
After the machine cleaning and disinfection,
a hot steam sterilization is required to dry the
breathing system. Valves, ventilator cover,
and AGS sleeve: Make sure that the cleaning
and rinsing liquids can flow in the opening
direction of the valve. Remove any water
which may have accumulated in the ventilator
diaphragm. For further instructions and
explanations please see the respective
instructions for use.
Visual check
Position components so that all interior spaces and surfaces are completely flushed and water can drain off freely.
This guideline does not replace the instruction for use.
D-58067-2012
Dräger Fabius Family
Compact Breathing System
Reprocessing Instructions