SupplementDaily Checkout
Daily Checkout
Heated Breathing System (when present)
Calibrate flow sensor.
Check cable connections.
Check power cord connections.
Ensure that the MAINS POWER LED is
illuminated.
Ensure that the HEATER POWER switch is in
the “ON” position.
Ensure that the “HEATER POWER ON” LED
is illuminated.
Ensure that the underside of the COSY is
warm (approximately 35-40°C) after 30
minutes have elapsed since the “HEATER
POWER ON” LED was illuminated.
Fabius GS/GS premi um/Tiro /OS Supplement Supplement #9038278, Rev. 07 17