Supplement General Warnings and Cautions
4 Supplement #9038278, Rev. 07 Fabius GS/GS premium/Tiro /OS Supplement
General Warnings and Cautions
The following list of warnings and cautions apply to
general operation and maintenance of the
Fabius. Warnings and cautions
about i
nstalling and operating specific parts appear
with those topics.
A Warning statement gives important
information that, if ignored, could lead directly
to personal injury.
A Caution statement gives important
information that, if ignored, could lead directly
to equipment damage and indirectly to
personal injury.
Warning: When the moisture content falls below a
specified minimum level, the following
undesirable reactions can occur,
regardless of the type of CO
2 absorbent
and the anesthetic agent used, e.g.
Halothane, Enflurane, Isoflurane,
Sevoflurane or Desflurane:
reduced CO
2 absorption,
formation of CO,
absorption and/or decomposition of
the inhalation anesthetic agent,
increased heat generation in the
absorber, leading to higher breathing
gas temperatures.
Additionally, the breakdown products of
anesthetic agents exposed to dry
absorbent are both flammable and toxic,
and fires have been reported in
association with the use of desiccated
absorbent and volatile anesthetics.
These reactions can result in danger to
the patient in the form of CO intoxication,
insufficient depth of anesthesia and
airway burns.

Other manuals for Dräger Fabius Tiro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Tiro and is the answer not in the manual?

Dräger Fabius Tiro Specifications

General IconGeneral
BrandDräger
ModelFabius Tiro
CategoryMedical Equipment
LanguageEnglish

Related product manuals