Technical data
336 Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Temperature
Parameter display Temperatures: Ta, Tb, ΔT, T1a, T1b, ΔT1
Measurement range Ta, Tb, T1a, T1b: 0 to 50 °C (32 to 122 °F)
ΔT, ΔT1: 0 to 50 °C (0 to 90 °F)
Resolution 0.1 °C (0.1 °F)
Accuracy (exclusive of probe) Ta, Tb, T1a, T1b: ±0.1 °C (±0.2 °F)
ΔT, ΔT1: ±0.2 °C (±0.4 °F)
Probe accuracy Reusable probe: ± 0.1 °C (±0.2 °F) at 0 °C to 50 °C (32 to 122
°F)
Disposable probes:
± 0.1 °C (±0.2 °F) at 25 °C to 45 °C (77 to 113 °F)
± 0.2 °C (±0.4 °F) at 0 to 25 °C (32 to 77 °F)
Average update time <2.5 s
Response time 23 to 44 °C (73.4 to 111.2 °F)
±0.2 °C (±0.4 °F) within 150 s
Degree of protection against electric
shock
Type CF
Defibrillation protection Yes