ALARM SETUP (ALARM LIMITS TABLE)
VF8 DELTA/DELTA XL/KAPPA 5-13
Quick Reference
– Alarm Limits Table Setup
Alarm Limits Table
Function Description Available Settings
Auto Set Sets alarm limits based on current values Not applicable
NOTES:
• The monitor recalculates the upper
and lower alarm limits based on the
parameter values in the Current
column.
• Auto Set applies to all displayed
parameters and ST parameters
only.
• If a calculated limit value falls
outside the range for that
parameter, the parameter’s alarm
limits will remain unchanged.
Alarms Enables or disables the alarm function for the
selected parameter
•ON
• (Alarm off icon)
Upper Determines current upper alarm limit Settings are parameter-specific
Current Read-only; cannot be modified Not applicable
Lower Determines current lower alarm limit Settings are parameter-specific
Parameters Upper Limit Lower Limit
Ta, T1a-b,
T2a-b, T3a-
b
< 107% of
current
value
< 93% of
current
value
T1, T2,
T3, PVC/
min, iCO
2
,
iCO
2
*
No change No change
SpO
2
/SpO
2
* Adults 100
Neonates
98
Current
value
- (value x
5%)
SpO
2
% Current
value + 20%
None
ST Current
value + 2.0
mm
Current
value
-2.0 mm
MultiGas
Agent
< 105% of
current
value
< 95% of
current
value
BIS No change No change
TOF-Cnt No change No change
PCCO Current
value + 30%
Current
value - 30%
PCCI Current
value + 30%
Current
value - 30%
MultiGas O
2
100% 21%
tp < 125% of
current
value
< 80% of
current
value