13-14 DELTA/DELTA XL/KAPPA VF8
13 N
ON-INVASIVE BLOOD PRESSURE
NBP Cuff Cannot
Deflate
Pneumatic failure. Check hose and cuff for obstructions
and replace, if necessary.
If the message does not clear,
contact the hospital’s technical
personnel or DrägerService.
NBP Cuff
Deflation Error
NBP pump or valves have been
energized for longer than 2 minutes
(Adult or Pediatric mode) or 90 seconds
(Neonatal mode).
Disconnect and reconnect cuff.
Check hose and cuff for obstructions
and replace, if necessary.
If the message does not clear,
contact the hospital’s technical
personnel or DrägerService.
NBP Cuff Leak The drop in cuff pressure after the end
of the inflation cycle is too great.
Check hose and cuff for leaks and
replace, if necessary.
Restart the measurement. If the
message does not clear, contact the
hospital’s technical personnel.
NBP
Cannot Measure
The pulse profile is too poor to make a
reliable measurement (usually due to
persistent motion artifact).
Check the patient and treat, if
necessary.
Move the cuff to a limb with less
movement.
Restart measurement. If the
message persists contact the
hospital’s technical personnel or
DrägerService.
NBP Blocked
Line
The inflation rate is too high during the
inflation cycle or the time to dump the
residual cuff pressure at the end of the
deflation cycle is too short.
Select a different cuff.
Check the hose and cuff for
damage.
Restart the measurement. If the
message does not clear and contact
the hospital’s technical personnel or
DrägerService.
NBP Cannot
Zero
The monitor is unable to zero
transducer within 30 seconds from start
of NBP program, usually because of
motion artifact.
Check the patient and treat if
necessary.
Move cuff to a limb with less
movement.
If the message does not clear,
contact the hospital’s technical
personnel.
NBP
Measurement
Timeout
A measurement was aborted (usually
due to motion artifact) because it lasted
longer than two minutes (adult or
pediatric) or 90 seconds (neonatal).
Repeat the measurement.
NBP No
Pulsation
Weak signal. The monitor is unable to
detect a sufficient number of pulsations
of adequate amplitude within two
minutes.
Check the patient and treat, if
necessary.
Check the hose and cuff.
Check for proper size, placement of
cuff.
Message Possible Cause Suggested Action