14-12 DELTA/DELTA XL/KAPPA VF8
14 I
NVASIVE BLOOD PRESSURE
Parameter boxes for non-pulsatile pressures (LA, RA, CVP, ICP, ICP2, ICP3, ICP4)
display mean pressures only:
The CPP parameter is calculated and displayed whenever the ICP and ART
parameters are monitored simultaneously (CPP = ART
mean
- ICP). CPP/CPP2/CPP3/
CPP4 is displayed in the bottom of the ICP/ICP2 /ICP3/ICP4 parameter box.
IBP waveforms can be displayed in standard or overlapping format. When IBP
waveforms are overlapped, you can show scale values for the overlapped parameters
side by side (the order of the display corresponding to the priority of the parameter
boxes) by setting the
Common Scale setting to OFF on the IBP setup menu (see page
14-14). The following figure shows overlapped IBP waveforms when the
Common
Scale
setting is set to ON.
Mean pressures
Parameter labels