STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 14-21
<xx>
Calibrating
Mercury calibration is in progress. Complete the calibration before you begin
monitoring the patient.
<xx> Cal.
Accepted
Mercury calibration succeeded or
the user-entered calibration factor
was accepted.
No action is required.
<xx> Cal.
Failed - Not
Static
The mercury calibration failed
because the input pressure was
not static.
Make sure the transducer is closed to
patient.
Check for leaks.
Keep all tubing motionless.
Rezero the transducer.
Refer to the calibration procedures (see
page 14-5).
Loosen and retighten the transducer dome,
then rezero the transducer.
Replace the transducer.
<xx> Cal.
Failed - Out of
Range
The mercury calibration failed
because the measured value was
too high or too low.
Make sure the transducer is zeroed, then
retry. If the retry fails, replace the transducer.
If calibration requires a factor outside this
range, replace the transducer.
<xx> Zero
before Cal.
During calibration, more than 5
minutes have elapsed since the
last successful zeroing of the
transducer.
Zero the transducer.
<xx> H/W
Failure
IBP channel hardware failure. Check the hardware and replace it, if
necessary.
Call the hospital’s technical personnel or
DrägerService.
Hemo Pod [n]
Disconnected
The hemo pod [1, 2, or 3] is not
connected to the monitor.
Check the cables and connections and
replace it, if necessary.
NOTE: <xx> represents the IBP parameter label associated with the displayed message.
Message Possible Cause Suggested Action