PHYSIOLOGICAL CALCULATIONS (HEMO/OXY/VENT CALCULATIONS)
VF8 DELTA/DELTA XL/KAPPA 16-9
Lab Data on Calculations Screen
To display laboratory data on the Hemo/Oxy/Vent Calculations screen, proceed as
follows:
1. Sample and analyze the patient’s blood.
2. Press the
Fast Access fixed key on the monitor’s front panel.
3. Click on
Calculations to display the Hemo/Oxy/Vent Calculations menu.
Lab data input parameters appear on the right.
4. Click on
Capture Values. A message at the top of the main screen indicates
when new lab data is available.
5. Click on
Capture Labs to capture blood gas values.
6. Click on
Calculate or Results. The Calculations screen appears with updated
derived parameters.
1
Time/date stamp indicating the time the
monitor receives data. Labeled
according to the following criteria: test
time – the time and date are generated
by blood analysis device and sent with
the lab results.
Arrival time – the time and date are
generated by the monitor; no test time is
sent with the lab results.
2
Displays current values (appears
ghosted until requested data is available)
3
A ‘greater than’ symbol (>) indicates that
the number of characters or digits
exceeds the available display space.
4
Parameter labels, values and units of
measure.