STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 17-13
SpO
2
Searching The monitor has not been able to
compute a valid measurement and is
searching for a pulse.
Make sure the SpO
2
sensor is
properly attached to the patient at a
site free of patient motion and all
cables are properly connected.
Make sure the correct sensor type
and/or site is used.
Contact Dräger DrägerService.
SpO
2
Unplugged The sensor cable is not connected to
the MultiMed pod.
Make sure the cables are securely
connected.
Check for a defective sensor.
SpO
2
: Pod Failure Pod hardware or software failure Disconnect/reconnect the pod, if the
problem persists.
Disconnect/reconnect the pod and/
or replace it with a good sensor.
Contact the hospital’s technical
personnel (SpO2 compatibility is a
locked option).
SpO
2
: Check/
replace Sensor
Defective/faulty sensor Replace with a Nellcor-compatible
sensor.
Contact the hospital’s technical
personnel (SpO2 compatibility is a
locked option).
SpO
2
: Connection
Not Supported
The Masimo SET pod is connected. Disconnect the Masimo SET pod and
connect a Nellcor Oximax pod.
Message Possible Cause Suggested Action