17-16 DELTA/DELTA XL/KAPPA VF8
17 P
ULSE OXIMETRY (SPO2)
Masimo SET pod
Message Possible Cause Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
is outside
the current upper or lower alarm
limits.
Check the patient and treat, if necessary.
PLS > #
PLS < #
The patient’s pulse rate is
outside the current upper or
lower alarm limits.
SpO
2
< Low limit
value - 20%
(Neonatal patient
category only)
The patient’s SpO
2
is below the
current lower alarm limit by 20%
or more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
Transparent Too much light is reaching the
sensor’s light detector. This
condition usually occurs
because the sensor is off the
finger or because too much
ambient light is detected.
Make sure that the sensor is properly
attached to the patient’s finger.
Remove the light source.
Cover the sensor with opaque material.
Make sure that no ambient light can
reach the detector.
Contact DrägerService.
SpO
2
Unrecognized
Sensor
The monitor does not recognize
the sensor connected as valid.
Check for a defective or unapproved
sensor.
Replace the sensor.
Contact the hospital’s technical
personnel or DrägerService.
SpO
2
Artifact A persistent artifact is detected. Make sure the SpO
2
sensor is properly
attached to the patient, patient is not
moving, and all cables are properly
connected.
Contact DrägerService.
SpO
2
Weak signal Pulse amplitude is too low due
to one of the following
conditions:
Poor perfusion (shock).
Low body temperature.
Check the patient and treat, if
necessary.
Make sure the SpO
2
sensor is properly
attached to the patient and all cables
are properly connected.
Relocate the sensor to another
extremity.
Contact DrägerService.
SpO
2
No
Measurement
The monitor has not been able
to compute a valid measurement
within the last 30 seconds
because of unstable
measurement conditions.
Make sure the SpO
2
sensor is properly
attached to the patient at a site free of
patient motion and all cables are
properly connected.
Contact DrägerService.