17-24 DELTA/DELTA XL/KAPPA VF8
17 P
ULSE OXIMETRY (SPO2)
SpO
2
* No
Measurement
The monitor has not been able to
compute a valid measurement
within the last 30 seconds
because of unstable
measurement conditions.
Make sure the SpO
2
sensor is properly
attached to the patient at a site free of
patient motion and all cables are
properly connected.
Contact DrägerService.
SpO
2
* Searching Searching for valid pulses from
which to compute
measurements.
Make sure the SpO
2
sensor is properly
attached to the patient.
SpO
2
* Regulation
Error
Inconsistent light level
Excess ambient light
Make sure the SpO
2
sensor is properly
attached to the patient.
Remove or shade any external sources
of light entering the sensor.
Contact DrägerService.
SpO
2
* Unplugged The sensor or sensor cable is
not connected to the MultiMed
pod.
Make sure that the cables and the
sensor are securely connected.
Check for defective sensor.
SpO
2
* Incompatible The MicrO2+ software version
does not match, or the MicrO2+
pulse oximeter is connected to a
Masimo-only or Nellcor-only
Monitor.
Contact DrägerService.
SpO
2
*
Disconnected
The MicrO2+ is disconnected. Reconnect the MicrO2+
SpO
2
* H/W Failure Front-end hardware circuit
failure.
Contact DrägerService.
Message Possible Cause Suggested Action