1-10 DELTA/DELTA XL/KAPPA VF8
1 I
NTRODUCTION
Monitor Right Side – Delta/Delta XL
Monitor Right Side – Kappa is blank
CAUTION:
The Kappa video display is not battery-backed. When power is lost, nothing
appears on screen unless the video display is connected to an Uninterruptable
Power Supply (UPS). Dräger recommends the use of a UPS with the Kappa
video display. The UPS must meet the electrical safety requirements of IEC
60601-1 or be connected to an isolation transformer that meets those
requirements (see the “Electrical Safety” section).
Connecting the Kappa monitor, Video display unit, and the optional UPS
requires multiple line cords. To reduce the chance of electromagnetic
interference from magnetic fields, the power cords should be run as close
together as reasonably possible to reduce loop area.
The video output connector on the back of the Kappa is not galvanically
isolated.
If you use a video monitor other than one specified by Dräger, it must comply
with IEC 60601-1 and be suitable for use in the presence of flammable
anesthetic mixtures (see “Safety Considerations” on page 6). Based on the
intended use of the system, the video monitor must also have suitable
classifications for water ingress protection as well as radiated and conducted
emissions. Upon installation, the installer must make sure that all of these
requirements are met.
1
PCMCIA slot: memory card
2
Infinity network connector
3
QRS sync
4
RS232 connector, Scio module, Smart Pod,
remote keypad, Vital Connect Cable (VCC),
Alarm output (nurse call)
1
– (X8)
5
Analog out (X10)
1
NOTE: Alarm output (nurse call) requires the
alarm output cable with partnumber 4314626.