STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 19-17
etCO
2
Out of
Range (High)
The CO
2
value is out of range
(high).
Check the patient and treat,if necessary.
Recalibrate sensor.
etCO
2
Check
Airway
Adapter/Cal
The airway adapter is dirty, not
fully seated, or out or calibration.
Make sure the adapter is properly seated.
Clean and calibrate the airway adapter.
etCO
2
Calibrating
Sensor
Calibrating on the zero cell in
progress.
Informational message; no action required
etCO
2
Cannot
Cal. Sensor
The calibration on the zero cell
could not be completed because
of CAPNOSTAT temperature
instability.
Check for any heat sources warming the
sensor and remove them.
Wait at least three minutes for the
temperature to stabilize.
etCO
2
Adapter
Failure
The airway adapter is dirty, not
fully seated, or out or calibration.
Make sure the airway adapter is properly
seated. Clean and calibrate the airway adapter
if needed.
etCO
2
Place
Sensor On
Ref Cell
The calibration on the zero cell
completed successfully.
Place sensor on the reference cell and wait for
calibration to complete.
etCO
2
Sensor
Cal. Failed
The calibration on the zero cell
failed.
Recalibrate. If the message persists, try a new
sensor.
etCO
2
Verifying
Sensor Cal
Calibrating on reference cell in
progress.
Informational message; no action required
etCO
2
Sensor
Cal. Verified
Verification completed
successfully.
Informational message; no action required
etCO
2
Calibrating
Adapter
The airway adapter calibration
(zeroing in room air) is in
progress.
Informational message; no action required
etCO
2
Cal.
Failed,
Breaths?
Breaths were detected during
the 20 second period following
the activation of the adapter cal.
key.
Make sure the sensor is not connected to the
patient’s ventilator breathing circuit or close to a
CO
2
source. Recalibrate.
etCO
2
Cannot
Cal. Adapter
The airway adapter calibration
(zeroing in room air) could not be
completed because of
CAPNOSTAT temperature
instability, or because the
CAPNOSTAT was on the zero cell.
Recalibrate holding the sensor in room air
(not on zero cell).
Wait at least eight minutes for temperature to
stabilize and recalibrate.
Remove any heat source warming the sensor
and recalibrate.
Remove sensor from Zero cell, place on the
adapter, and recalibrate.
etCO
2
Adapter
Cal. Accepted
The airway adapter calibration
(zeroing in room air) is
completed.
Informational message; no action required
Message Condition Suggested Action