SCIO S ETUP
VF8 DELTA/DELTA XL/KAPPA 23-15
O
2
/N
2
O Monitoring (Scio Four Oxi plus and Scio Four Oxi
only)
The O
2
waveform indicates O
2
concentrations calculated by the Scio module. The O
2
parameter box can display the current concentration values for the following
parameters:
Inspired O
2
(iO
2
)The level of O
2
in the airway during the inspiration phase
Expired O
2
(etO
2
) The level of O
2
in the airway during the expiration
phase
N
2
OThe concentration of N
2
O in the patient’s airway
The appearance of the O
2
parameter box varies depending on whether or not the N
2
O
display is turned on in the O
2
menu.
Typical O
2
/N
2
O parameter box displays are shown below.
To access the O
2
(/N
2
O) setup menu
Click on the O
2
/N
2
O parameter box
or
1. Press the
Menu fixed key.
2. Click on
Patient Setup.
3. Click on
Parameters to display a list of available parameters.
4. Click on
O2.
Parameter Box Description
Scio module:
N
2
O display turned off.
This module calculates both iO
2
and etO
2
values.
Scio Module:
N
2
O display turned on.
This module calculates both iO
2
/iNO
2
and etO
2
/etNO
2
values.
NOTES: The 1 symbol in the parameter box indicates that the O
2
lower alarm limit
has been set to a value less than 21% (the percentage of O
2
in room air).
The O
2
parameter box does not show N
2
O alarm limits because N
2
O does not alarm.