26-22 DELTA/DELTA XL/KAPPA VF8
26 P
ULSE CONTOUR CARDIAC OUTPUT (PICCO) MONITORING
CVP Value Allows user to select a CVP value.
NOTE: Auto CVP Value, if available,
is used until the manual CVP Value
menu is enabled. If auto CVP Value
is unavailable, a default of 5mmHg is
used until manual CVP Value is
menu is enabled.
5 mmHg (default)
0 mmHg to 300 mmHg
(in increments of 1 mmHg)
Injectate Volume Selects the injectate volume. 15.0 cc (default),
2.0 cc to 20 cc, in increments of
1.0 cc
p-CO Start Brings up a p-CO Averaging screen
and starts p-CO measurement.
NOTES:
• After each thermodilution
measurement completes, select
p-CO Start in the Averaging screen
to perform another measurement
(the monitor does not automatically
start a new measurement).
• A p-CO measurement can be
started using the CO Start key on
the PiCCO pod (the keypad's C.O.
key does not work).
• If p-BT or p-IT values are invalid or
out of range, the p-CO Start button
is ghosted and the Averaging
screen cannot be accessed. During
this time the C.O. Start fixed key on
the PiCCO pod is also inactive.
Not applicable
p-BT <value>
p-IT <value>
Read only display of p-BT and p-IT
values.
NOTE: If p-BT or p-IT values are out
of range, the values will display “***”.
p-BT range: 25 to 43°C (77 to 109°F)
p-IT range: -3 to 31°C (27 to 88°F)
Not applicable
PBW <value>
PBSA <value>
Read only display of PBW and PBSA
values.
Not applicable
Manual CVP Log CVP value is logged in with a time
stamp.
Not applicable
NOTE: A total of 20 entries per
patient can be stored in the log.
Thermodilution Setup Menu
Menu Item Description Available Settings