SYSTEM C OMPONENTS
VF8 DELTA/DELTA XL/KAPPA B-11
Kappa CPU Base Unit
Physical Attributes
Size (H x W x D): 102 x 368 x 368 mm (4 x 14.5 x 14.5 in)
Weight: 8.4 kg (19 lb.)
Cooling: Fan
Enclosure: Plastics: ABS/PC, FR 110
Painted, cold-rolled steel
Printed circuit boards: Board: Glass/epoxy Fr4
Solder: Lead/tin
Copper etch
Lithium battery
Heatsink assemblies: Cast aluminum
NBP assembly: Silicon tubing, steel, copper wire
Electrical Specifications
Power requirements: 100-240 VAC, 2.5 A to1.3 A 50/60 Hz
Protection class: Internally powered (per IEC 60601-1) and for use with specified
Class 1 power supplies.
Chassis leakage: 300 A @ 120 VAC
500 A @ 240 VAC
Battery type: Lithium Ion
Battery charge duration: 20 minutes
Battery charging time: 3.5 hours at 25 ºC
Patient leakage current: 10 A
Mode of operation: Continuous with external power supply; for a limited time with
battery backup
NOTES:
• Battery life varies with parameter configuration. The battery life specified above is under the
following load conditions: MultiMed
with SpO
2
sensor, 2 temperature probes, HemoMed pod with
4 IBP transducers and a catheter, NBP taking measurements every 15 minutes, LCD Transport
Brightness at 50%, and no continuous tone being generated.
• Advanced Communication Options are not supported during battery operation.
Environmental Requirements
Temperature range: Operating: 10 to 45 °C (50 to 113 °F)
Storage: –15 to +50 °C (–5 to +122 °F)
Relative humidity: Operating: 20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure: Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Protection against ingress of
water:
IPX1
Analog Output
Signals: ECG, arterial blood pressure (ART)