B-16 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
R50 N Infinity Recorder
Physical Attributes
Size (H x W x D):
180 x 120 x 222 mm (7.1 x 4.72 x 8.74 in)
Weight:
1.64 kg (3.6 lb)
Connections:
AC power connector; X14 Infinity Network; X7 R50 recorder;
potential equalization connector
Cooling:
Convection
Type:
Transportable equipment
Electrical Specifications
Input voltage range:
100 to 240 VRMS
Mains frequency:
50/60 Hz
Power consumption:
1.0 A max
Protection class:
Class 1
Chassis leakage current:
<300 A @ 120 VAC, <500 A @ 220 VAC
Mode of operation:
Continuous
Protection against water ingress:
Ordinary (IPX0 per IEC 60529)
Fuses
Replace as marked, F2A-250V
NOTE: There are no other user-replaceable parts for this device.
Environmental Requirements
Temperature range:
Operating: 15 to 40 °C (55 to 104 °F)
Storage: –20 to 40 °C (–4 to 104 °F)
Relative humidity:
Operating: 30 to 95 %, non-condensing
Storage: 10 to 95 %, non-condensing
(with packaging)
Atmospheric pressure:
Operating: 550 to 775 mmHg (73 to 103 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)