MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-27
S
CIO Four/SCIO Four Oxi/SCIO Four plus/SCIO Four Oxi plus
Module
Physical Attributes
Size (H x W x D) with watertrap: 115 x 190 x 270 mm (4.8 x 7.5 x 10.5 in)
Weight: Scio Four:3.024 Kg (6.66 lbs)
Scio Four Oxi:3.444 kg (7.59 lbs)
Scio Four plus:3.037 kg (6.69 lbs)
Scio Four Oxi plus:3.457 kg (7.62 lbs)
Cooling: Fan
Mains frequency: 50/60 Hz
Power requirement: <0.8 A at 100 to 120 VAC; <0.4 A at 200 to 240 VAC
Chassis leakage current: 300 A (per UL 544)
A (per IEC 60601-1)
Electric shock protection: Type BF
Protection class: Class 1
Mode of operation: Continuous
Power: from specified power supply
Sound pressure level 45 dB(A)
Air ingression, leakage <45 ml during zeroing, <10 mL/min leakage
Sample flow rate 150 ±20 mL/min
(or 200 ±20 mL/min, if so specified on back-panel)
Environmental Requirements
Temperature range: Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 70 °C (–4 to +158 °F)
Relative humidity: Operating:5 to 90 %
Storage: 5 to 95 %
Atmospheric pressure Operating:525 to 795.1 mmHg (70 to 106 kPa)
Storage: 375 to 795.1 mmHg (50 to 106 kPa)
NOTES:
• Readings comply to ATPS conditions.
• This device is not intended for use in the presence of flammable gases.
Regulatory Requirements
Standards: IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable
sections only, Oxygen analyzers for monitoring patient
breathing mixtures); EN ISO 21647:2009