2-16 DELTA/DELTA XL/KAPPA VF8
2 M
ONITOR SETUP
NBP/SpO2
Interlock
Allows user to render SpO
2
alarm inactive
when NBP measurement is in progress.
WARNING: Visually check that
the NBP cuff is on the same arm
as the SpO
2
sensor. The monitor
will not automatically detect
that NBP cuff and SpO
2
sensor
are on the same arm.
• ON – The SpO
2
alarm function
is inactive during an NBP
measurement.
• OFF (default) – The SpO
2
alarm
is active during an NBP
measurement.
Alarm Limits
Display
Enables/disables the display of alarm
limits in the parameter boxes.
• ON (default) – Alarm limits
display next to the current
parameter value in the
parameter box. A crossed bell is
displayed there if that alarm is
disabled.
• OFF – Alarm limits are not
displayed.
Alarm Light Allows you to enable/disable the alarm
bar which is located on top of the
monitor’s housing.
NOTE: The alarm bar is not available on a
SC 7000/8000/9000XL or Kappa monitor.
NOTE: During multiple alarm conditions,
the alarm bar flashes only for the alarm
condition with the highest priority.
• ON (default) – Alarm bar reacts
as follows:
- Flashes red for all high-priority
alarm conditions.
- Flashes yellow for all medium-
priority alarm conditions.
• OFF – The alarm bar is
disabled.
MIB Alarm
Control
Allows you to activate/deactivate MIB
disconnect alarms associated with
ventilators and/or anesthesia machines.
NOTE: This selection applies only with
ventilators and/or anesthesia machines
• ON (default) – MIB disconnect
alarms are active.
• OFF – MIB disconnect alarms
are inactive.
Remote View
Display
Allows user to set Remote View (see
page 3-13) behavior when the Remote
View is displaying telemetry ECG data.
• Always ON – the local bed
alarm will not cause remote
view feature to “pull down” the
remote bed display.
• Pull Down on Alarm (default) –
the local bed alarm will cause
remote the view feature to “pull
down” remote bed display.
The Unit Manager Menu
Menu Item Description Available Settings