2-22 DELTA/DELTA XL/KAPPA VF8
2 M
ONITOR SETUP
Biomed functions are described in the following table:
Quick Reference – Biomed Menu
Menu Item Description Settings/Procedures
The
Logs Submenu
This menu displays clinical and technical diagnostic records. Open the Biomed menu (page 2-21)
and click on Logs, then follow the procedures described in this table.
Component Log The monitor maintains read-only
logs for principal components or
devices. These logs include a part
number, revision numbers, serial
numbers, software version, and
compatibility information.
1) Scroll to the component you
wish to inspect and click on it.
The log appears with the exit
arrow already selected.
2) Return to the Logs submenu by
clicking again.
Status Log Displays information about your
current software and hardware
versions.
The display is read-only.
Diagnostic Log Captures data about hardware and
software performance relating to the
monitor’s operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
CPS/IDS Diag. Log Captures data about CPS and IDS
performance relating to the monitor’s
operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
FE Diag. Log Captures and displays data about
front-end performance related to the
monitor’s operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
Copy All Logs Downloads status logs and
diagnostic logs to a memory card.
Click on Copy All Logs. A
confirmation message appears in
the message area to indicate the
download is complete.
Print Log Prints an expanded version of the
Diagnostic Log to an Infinity network
laser printer
Click on Print Log.
The
Service submenu
The Service menu is password-protected
and intended for the hospital’s technical personnel or DrägerService personnel only.
Test Pulse
Test Pulse Tests the ECG and EEG signal
clarity and display.
Clicking on Test Pulse causes the
following to happen:
• Injects a 300 ms pulse into the
ECG waveform (1 mV on leads I
and III, 2 mV on lead II).
• Superimposes a 100 µV pulse on
all EEG channels for 200 ms.