INFINITY DELTA SERIES USER’S GUIDE
VF8 DELTA/DELTA XL/KAPPA 9
Before docking, undocking or moving a monitor, verify that the mounting mechanism
is mechanically sound. Be careful not to apply too much force when docking the
monitor.
Verify that the safety labels are legible and the safety checks were performed at the
required interval.
Safety checks, verification, calibration and maintenance should be performed at least
every two years by properly trained personnel, as described in the Service manual (see
individual parameter chapters for information about calibration and verification of
parameter-specific functions and devices). All cables, alarm functions, accessories,
and associated devices should be checked for damage, ground resistance, chassis and
patient leakage currents on a yearly basis, or more frequently based on usage.
Maintain a record of these safety checks and other inspections.
WARNING:
Repair of the device may only be carried out by trained
service personnel otherwise the correct functioning of
the device may be compromised. Regular annual
maintenance (functional and safety test) according to
IEC 62353 is recommended, in addition to national
regulations and laws (for example, accident prevention
regulations). Connecting this medical device to other
medical devices could result in additional maintenance
requirements. Consult the documentation for these
other devices to identify additional requirements.
Dräger recommends contracting with DrägerService for
any repairs. Use only authentic Dräger repair parts
during maintenance. Using non-Dräger repair parts may
adversely affect the operation of the device.
Contact your hospital’s technical personnel if the
monitor’s mounting mechanism appears mechanically
damaged or its structural integrity is compromised. Do
not mount the monitor under such circumstances.
NOTE:
The monitor’s service manual is available from your local DrägerService repre-
sentative.
Dispose of all equipment in accordance with local regulations.