Pocket Guide Oxylog 3000 plus 13
Assembly
Connecting the adult disposable
hose system
1 Connect the flow measuring hoses (A) to the
Oxylog 3000 plus. Correct alignment is indicated by
a notch on the connector, which must point away
from the ventilation hose. Otherwise, the set will not
fit and the measured values will be incorrect.
2 Connect the ventilation hose (B) to the gas outlet on
the Oxylog 3000 plus.
When connecting a hose, check that the hose setting in
the Settings window corresponds to the connected ho-
se.
Connecting the paediatric
disposable hose system
1 Connect the blue flow measuring hose (B) to the
blue labeled connector.
2 Connect the transparent flow measuring hose (A) to
the other connector.
3 Connect the ventilation hose (C) to the gas outlet on
the Oxylog 3000 plus.
When connecting a hose, check that the hose setting in
the Settings window corresponds to the connected ho-
se.
013
A
B
010
B
C
A