Operation
Operation
Ventilation functions of the
Oxylog 3000 plus
Ventilation modes:
Volume-controlled ventilation:
VC-CMV / VC-AC
–VC-SIMV
Pressure-controlled ventilation:
–PC-BIPAP
Support of spontaneous breathing:
SpnCPAP
Additional settings for ventilation:
Pressure support: in the ventilation modes
VC-SIMV, PC-BIPAP, SpnCPAP
Apnoea ventilation: in the ventilation mode
SpnCPAP
AutoFlow (optional): in the ventilation modes
VC-CMV, VC-AC, and VC-SIMV
NIV: in the ventilation modes SpnCPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF, and
VC-SIMV / AF
Special procedures:
Inspiration hold
–O
2 inhalation (optional), with an inhalation mask
VC-CMV, VC-AC
Volume Controlled - Controlled Mandatory
Ventilation
Volume-controlled ventilation with fixed mandatory mi-
nute volume MV, which is set with tidal volume VT and
respiratory rate RR.
Set the ventilation pattern with the controls below the
display:
Tidal volume VT.
Ventilation respiratory rate RR.
(minimum possible frequency: 5 per min).
Maximum airway pressure Pmax.
–O
2 concentration FiO2.
NOTE
The Oxylog 3000 plus Pocket Guide is not a replacement or substitute for the Instructions for Use.
Any use of the device requires full understanding and strict observation of the Instructions for Use.
503
Paw
Flow
T
insp
T
e
P
plat
Plateau time
Tplat
1
RR
Insp. Flow
t
t
P
max
PEEP
Exp. Flow