6
11.1 Trouble shooting errors
Error
code
Cause Remedies
010 "Horn alarm elements" X-
dock test failed
Repair if necessary and
test again with X-dock
011 "LED alarm elements" X-
dock test failed
Repair if necessary and
test again with X-dock
012 "Vibration alarm elements"
X-dock test failed
Repair if necessary and
test again with X-dock
013 Parameter check failed Correct parameters and
test again with X-dock.
014 Instrument locked by X-
dock
Cancel X-dock lock
100 Flash/EEprom write failed Contact Service
104 Flash check sum wrong Contact Service
105 Broken or missing O
2
sensor
Replace O
2
sensor
106 Most recent settings
restored
Check settings and
recalibrate instrument
107 Self test failed Contact Service
108 Configuration not up-to-
date
Reconfigure with the latest
Dräger CC-Vision
software version
109 Configuration error Reconfigure instrument
161 Set operating time of the
instrument expired
Set new operating time of
the instrument
210 Zero-point/fresh air
calibration failed
Perform zero-point/fresh
air calibration
220 Span calibration failed Perform span calibration
221 Calibration interval
expired
Calibrate instrument
240 Bump test failed Perform bump test or
calibration
241 Bump test interval expired Perform bump test or
calibration
Warnin
g code
Cause Remedies
162 Set operating time of the
instrument almost expired
Set new operating time of
the instrument
222 Calibration interval
expired
Calibrate instrument
242 Bump test interval expired Perform bump test or
calibration
12 Technical Specifications
12.1 General
12.2 Standard Configuration (Factory Settings)
Environmental Conditions
During operation temperature see see chapter 12.3 and
see chapter 12.4
700 to 1300 hPa
10 to 90 % relative humidity
Conditions for
storage
0 to 40
o
C 32 to 104
o
F
30 to 80 % relative humidity
Battery life
(typical at 25
o
C)
24 hours of use per day,
1 minute alarm per day:
>5,500 hours, O
2
: >2,700 hours
Intensity of alarm typical 90 dBA at 30 cm/1 ft.
Dimensions
(without clip)
64 x 84 x 20 mm
(battery compartment
25 mm)
2.5 x 3.3 x 0.8 (battery compartment 1 in.)
Weight 106 g/3.8 oz.
Ingress protection IP 68
Approvals (see “Notes on Approval” on page 267)
Bump test mode
1)
1 May differ for customer-specific orders.
Quick bump test
Vibration alarm Yes
Bump test interval
1)
Off
Life signal
1)
2)
2 For measurements according to EN 45544 (CO, H
2
S) or
EN 50104 (O
2
) the life signal must be switched on.
Off
Turning the instrument off
1)
Always allowed
Data logger interval 1 minute
Operation timer Off
% COHB mode Off