18 Instructions for use VarioLux
Classification of device-specific components
The following classification is a recommendation from Dräger.
Non-critical
Device surface
Device-specific components
Refer to the "Overview" chapter on page 15.
Before reprocessing
Observe before disassembly
1 Turn off the device with the on/off switch (F).
2 Disconnect the power cable (E) from the mains connection
cable (D).
Validated reprocessing procedures
Overview of the reprocessing procedures of the
components
Surface disinfection with cleaning
Prerequisites:
The surface disinfectant has been prepared in accordance
with the manufacturer's instructions.
The manufacturer's instructions, e.g., regarding shelf life or
application conditions, are observed.
An uncontaminated, lint-free cloth soaked in surface
disinfectant is used for the cleaning surface disinfection.
Cleaning
1 Wipe off obvious soiling with a disposable cloth soaked in
surface disinfectant. Dispose of the cloth.
2 Wipe all surfaces. After that, there must no longer be any
soiling visible.
Surface disinfection
3 Wipe cleaned surfaces again to visibly wet all surfaces to be
disinfected with surface disinfectant.
4 Wait for the surface disinfectant contact time.
5 At the end of the contact time, moisten a new, uncontaminated
and lint-free cloth with water (at least drinking water quality).
6 Wipe all surfaces until no remains of the surface disinfectant,
such as foam residues or streaks, are visible.
7 Wait until the surfaces are dry.
8 Check the surfaces for visible damage and, if necessary,
replace the product.
Other agents and reprocessing
procedures
Disinfectants
Use nationally approved disinfectants suitable for the respective
reprocessing process and the intended application.
Surface disinfectant
The manufacturers of the surface disinfectants have verified at
least the following spectra of activity:
Bactericidal
Yeasticidal
Virucidal or virucidal against enveloped viruses
Follow the manufacturer's instructions for surface disinfectants.
Components Surface disinfection with cleaning
Device surface Yes,
refer to "Surface disinfection with clean-
ing"
Surface dis-
infectant
Manufacturer Concentra-
tion
Contact
time
Dismozon
pur/plus
BODE Chemie 1.6 % 15 min

Questions and Answers: