PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 2
Assembly of the Resuscitaire Birthing Room Warmer (WBR82)
CAUTION:
When removing the equipment from the carton, use care not to scratch or otherwise damage
unprotected surfaces.
Warmer
1. If a monitor shelf is to be installed on the unit, install the
weights provided with the monitor shelf in the bottom of the
lower post. Refer to the instructions provided with the monitor
shelf.
2. Remove the two #8-32 x ¾" screws that secure the controller
to the head support plate. Retain the two #8-32 x ¾" screws.
Slide the controller out through the front of the upper post.
WARNING:
Use two persons of sufficient strength to lift the upper
post. Failure to do so could result in personal injury or
equipment damage.
3. Using two persons of sufficient strength, lift the upper
post, and rest it on top of the opening of the base
weldment.
4. Fully extend the suction hoses from the upper post and
the base weldment, and then connect them to each
other.
5. Carefully feed excess hose into the upper post.
6. Release the suction hoses from the Neat-clip located
inside the base weldment.
7. If necessary, reposition the upper post on the base
weldment assembly.
8. Install the four #10-32 x 3/8" Nylok screws provided to
secure the upper post to the bassinet/cart assembly.
CAUTION:
Do not kink the suction hoses. Equipment
damage could occur.
9. Being careful not to kink the suction hoses, push them
into the upper post.