PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 6
Assembly of the Resuscitaire Radiant Warmer (RW82)
NOTE:
The required mounting hardware is stored in a bag in the pass-through drawer.
Warmer
1. Remove the two #8-32 x ¾" screws that secure the controller
to the head support plate. Retain the two #8-32 x ¾" screws.
2. Slide the controller out through the front of the upper post.
WARNING:
Use two persons of sufficient strength to lift the upper
post. Failure to do so could result in personal injury or
equipment damage.
3. Using two persons of sufficient strength, lift the upper
post, and rest it on top of the opening in the column of
the bassinet/cart assembly.
4. Fully extend the suction hoses from the upper post and
the bassinet/cart assembly, and connect them to each
other.
5. Feed excess hose into the upper post.
6. Release the suction hose from the Neat-clip inside the
base weldment.
7. If necessary, re-position the upper post on the
bassinet/cart assembly.
8. Install the four #10-32 x 3/8" Nylok screws provided to
secure the upper post to the bassinet/cart assembly.
CAUTION:
Do not kink the suction hoses. Equipment
damage could occur.
9. Being careful not to kink the suction hoses, push them
into the upper post.