PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
5 - 11
Name Description
Suction gauge Displays the suction level from 0 mm Hg (0"H
2
O/mbar) to 200
mm Hg (107"H
2
O/mbar) of vacuum.
Suction control Adjusts the suction level from 0 mm Hg (0"H
2
O/mbar) to 150
mm Hg (80"H
2
O/mbar).
Suction On/Off switch Turns the suction on and off. (Gas supply must be turned on.)
Rate (BPM) control (AutoBreath™
Infant Resuscitator Option)
Adjusts breath frequency from 18 to 60 breaths per minute.
PEEP control (AutoBreath™
Infant Resuscitator Option)
Adjusts positive end expiratory pressure from 0 to 1.8 kPa.
On/Off switch (AutoBreath™
Infant Resuscitator Option)
Turns AutoBreath™ Infant Resuscitator on and off (including
PEEP).
Exhalation valve (AutoBreath™
Infant Resuscitator Option)
Accepts exhaust valve line of patient circuit for expiratory valve
control.
Airway Pressure gauge Displays the airway pressure from -1 kPa (-10 cm H
2
O/mbar) to 8
kPa (81 cm H
2
O/mbar).
Airway Pressure Relief control Adjusts the airway pressure relief setting from 0 kPa
(0 cm H
2
O/mbar) to 5 kPa (51 cm H
2
O/mbar).
Flow Rate (LPM) control (flow
meter)
Adjusts the patient gas flow from 0 LPM to 15 LPM.
Patient Outlet connector Accepts the patient circuit outlet line.
Accepts the breathing circuit (PN 81 001 27/ MU10900or PN 81
000 06/ MU10841). If the blender option is incorporated, it deliv-
ers the blended gas.
Airway pressure port Connects the airway pressure gauge to the patient breathing
circuit.
Auxiliary Outlet 100% oxygen
connector
Connection for the auxiliary gas delivery line at the hose barb
connection on the front panel. Provides 100% oxygen only. Use
tubing with an inner diameter of 0.25" (0.64 cm).
Auxiliary Flow (LPM) control Adjusts the auxiliary patient gas flow from 0 LPM to 15 LPM.