TROUBLESHOOTING
Note: The latch must be turned completely clockwise to its stop position
in order for the centrifuge to operate.
The rotor does not spin
freely.
o Make sure nothing has fallen into
the rotor chamber.
o If there is nothing obstructing the
rotor, the rotor may be damaged.
contact Drucker Diagnostics for
further assistance.
Excessive noise when
the machine is
running.
o Check to see that the load is
balanced.
o Make sure that nothing has fallen
into the rotor chamber.
o Make sure that the nut in the
center of the rotor is tight.
o Have a technician test the motor
and replace it if necessary.
The centrifuge does
not run.
o Check the electrical outlet.
o Make sure the lid latch is turned
completely clockwise to its stop
position.
o Check the circuit breaker switch at
the bottom left of the machine. If
the switch is white, the breaker has
tripped. Contact Drucker
Diagnostics for further assistance.
o The printed circuit board may be
damaged. Have a technician test
and replace the circuit board if
necessary.