QBC STAR System Operator’s Manual
6000-300-000 2–4
Reporting Ranges
Parameter Lower Limit Upper Limit
Hematocrit (Hct) 15% 65%
Hemoglobin (Hgb) 5.0 g/dL 20.0 g/dL
MCHC 25.0 g/dL 37.3 g/dL
White Blood Cells (WBC) 1.6 x 10
9
/L 99.9 x 10
9
/L
Granulocytes 0.8 x 10
9
/L 70.0 x 10
9
/L
% Granulocytes 1% 99%
Lymph/Mono 0.8 x 10
9
/L 99.9 x 10
9
/L
% Lymph/Mono 1% 99%
Platelets 20 x 10
9
/L 999 x 10
9
/L
External Connections
The following external connections are located on the rear of the QBC STAR instrument, and are shown in Figure 4:
DC power input
USB Port (3x) for external printer, keyboard, USB memory stick, or scanner
Ethernet Port
You can attach a keyboard, USB memory stick, barcode scanner, and a full-size external printer (for print-
ing results on standard letter or A4 size paper) via the USB ports in the back of the QBC STAR instrument.
You can also leave the internal instrument printer enabled even if you attach an external printer. If you do,
results print on the internal printer and the external printer at the same time.
For information on what makes and models of printers can be used, contact Drucker Diagnostics Technical
Support or visit www.druckerdiagnostics.com.