XS series operating manual 
 
XS user manual v0.55  - 125 - 
BEEP ON/OFF  
This command is attached to the configuration function and controls the beep signal when 
the measurement result is false. 
It is not possible to stop the beep signal completely. 
For instance : CONF:DISP OFF 
 
DISP(DISPlay) ON/OFF  
This command is attached to the configuration function and controls the display mode. 
For instance : CONF:DISP OFF 
 
CAD (CApaDet) ON / OFF 
 
This command is attached to the configuration of the SXS16 hipot function (hipot 10VA). It 
activates/deactivates the detection of capacitive samples. 
             Ex : HIP:CAD ON 
 
[expert] 
EXP(EXPert) ON / OFF 
Set the EXPERT mode ON or OFF  
[expert] 
M335 ON / OFF 
Set the mode power in accordance with the standards 60335 et 60204. 
 
[expert] 
 
CTH (CapaTH) <NR1> 
 
This  command  sets  the  capacitive  value  threshold,  in  nF,  to  detect,  when  the  detection  of 
capacitive samples is activated.  
 
MOD (MODe) AUT/MAN 
This  command  is  attached  to  the  setup  function.  It  switches  the  mode  AUTO  and 
MANUAL of the sequence function (only with the family units of SMG and FMG). 
 
SBS (Step By Step) ON/OFF 
This command takes part of sequence function of the SMG and FMG. 
It fixes on the way of the test results are send back to the controller (RS232, IEEE). 
In ON mode, the result of every steps of test is transmitted as soon as available, without 
any request before (no MEAS?). 
In OFF mode, all the results are transmitted together at the end of the sequence in answer 
to the command MEAS? 
IMPORTANT : After power on the mode SBS is automatically set with OFF value. 
 
 
TIM(TIMe) AUT/FAIL/UDIV2   
Select  the  temporization  mode.  AUT  for  automatic,  FAIL  for  default  and  UDIV2  for  U 
divided by 2. 
The UDIV2 mode concerns the hipot function. 
The  FAIL  mode  concerns  the  ground  continuity  function  and  the  hipot  500VA function 
only. 
 
FILT(FILTer) NOR/CAP/RHT   
Enables the normal or the capacitor or the Real Hold Time measurement modes. 
For instance : CONF:FILT CAP