HEIGHT REGULATION
3.8” to 4.7” / 97.5 to 120 mm
WEIGHT
46.3 pounds / 21 kg
SERVICE CONDITIONS
TEMPERATURE
50 104 ºF
HUMIDITY
35 - 75 % relative humidity
PRESSURE
700 - 1060 hPa
STORAGE AND TRANSPORTATION CONDITIONS
TEMPERATURE
-22 122 ºF
HUMIDITY
10 - 95 % relative humidity
PRESSURE
500 - 1060 hPa
9. TROUBLESHOOTING
The following table indicates the most common problems and their quick solutions. Please consult
this table for any difficulty you may encounter.
PROBLEM
PROBABLE CAUSE
ACTION
The light source does
not work
It is not connected to the
local electrical network
The AC interlock cable is
broken or cut.
The plug is not in good
conditions
The fuse is broken or
burnt out
Network’s voltage is not
compatible with the one
required
Connect the cable to the network
Change the cable
Change the plug
Change the fuse
Verify voltages
There is no illumination
but the green pilot light
is on
The lamp is burnt
Select the other lamp or change
the burnt one
The intensity of the light
is not correct
The lamp is damaged
The lamp is out of place
Network’s voltage is 110
V but the voltage selector
indicates 220 V
Replace the lamp
Verify that the lamp is secured in
place
Select 110 V
High temperature at the
light source
The fans do not work
Obstruction of the
Notify the technical service
Liberate or clean the ventilation