EC9830 CO ANALYZER OPERATION MANUAL
2-4 98307600 Rev. C-1
Figure 2-3. 50-Pin Connector Board with Sample Choices (Rear)
Select the output for your application using the following steps.
1. Remove the 50-pin connector board from the rear panel of the analyzer.
2. Place the jumpers on the pins that correspond to the desired printed selections
on the front of the board. If current is selected, only the jumpers selecting
current make contact with both rows of pins. The other jumpers are offset as
shown in Figure 2-3.
If a current output is selected, the range must also be chosen from the
menu when the instrument is operating. The compliance voltage for the
current output is 12 V. A terminating resistor of 600 ohms or less should
be used for measurement errors no greater than 1%.