EasyManua.ls Logo

EDAN F6 - Page 176

EDAN F6
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
F Series Fetal & Maternal Monitor User Manual Ultrasound Intensity and Safety
- 165 -
Operating Mode: PW mode
Working Frequency: 1.0MHz
Acoustic Output MI
I
spta.3
(mW/cm
2
)
I
sppa.3
(W/cm
2
)
Global Maximum Value 0.01567 1.817 0.01025
Associated Acoustic
Parameter
P
r.3
(MPa) 0.01567
W
0
(mW) 11.52 11.52
f
c
(MHz) 1.00 1.00 1.00
Z
sp
(cm) 13.65 13.65 13.65
Beam
dimensions
X
-6
(cm) 0.5795 0.5795
Y
-6
(cm) 0.6061 0.6061
PD (usec) 88.72 88.72
PRF (Hz) 2000 2000
EDB
A
z
(cm) 3.46
E
le
(cm) 3.46
Operating Control
Conditions
Fixed

Table of Contents