EasyManua.ls Logo

EDAN F6 - A4.6 Probe Acoustic Output Parameters List; A4.6.1 Test of Wired Probe

EDAN F6
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
F Series Fetal & Maternal Monitor User Manual Ultrasound Intensity and Safety
- 164 -
A4.6 Probe Acoustic Output Parameters List
A4.6.1 Test of Wired Probe
Acoustic Output Reporting Table (F9/F9 Express)
Operating Mode: PW mode
Working Frequency: 1.0MHz
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-scan
Aaprt1 Aaprt>1
Global Maximum Index Value 0.01567 0.008761 0.05723 N/A
Associated
Acoustic
Parameters
P
r.a
(MPa) 0.01567
P (mW) 11.52 N/A
Min of [P
a
(Z
s
),I
ta.a
(Z
s
)] (mW) 1.84
Z
s
(cm) 13.95
Z
bp
(cm) 5.188
Z
b
(cm) 13.60
Z at max I
pi.a
(cm) 13.80 13.80
deq(Z
b
) (cm) 1.75
f
awf
(MHz) 1.00 1.00 1.00 N/A
Dim of Aaprt
X (cm)
3.46 3.46 N/A
Y (cm) 3.46 3.46 N/A
Other
Information
td (usec) 88.72
prr (Hz) 2000
P
r
at max I
pi
(MPa) 0.02930
Deq at max I
pi
(cm) 1.73
I
pi.3
at max MI (W/cm
2
) 0.01025
Focal
Length
Flx (cm)
N/A
Fly (cm) N/A
Operating
Control
Conditions
Focus(mm) Fixed
Depth(mm) Fixed
Freq(MHz) 1.0

Table of Contents