Patient Monitor User Manual Product Specification
- 243 -
ESU equipment
The wireless network function of all monitors
works normally and meets the following
requirements:
Total delay time for data transmission from the
monitors to MFM-CMS: 2 s.
Total delay time of data transmission from one
monitor to other monitors: 2 seconds.
Effective time of alarm reset configured on
another monitor 2 s.
Effective time for monitor-related settings
configured on the MFM-CMS: 2 s.
No communication loss between all the monitors.
Wi-Fi Network Stability
When the following conditions are present,
Quantity of the monitors supported by a single
AP: 8.
Each monitor can communicate with MFM-CMS.
Each monitor supports bed view function, which
allows users to view its information from another
bed or view other bed’s information from its screen.
The AP signal strength of the monitor should be
stronger than -65 dBm.
The following requirements must be met:
Within 24 hours, the time percentage of failing to
transmit data from any monitor to the MFM-CMS
does not exceed 0.1%. When the connected 8
monitors roam for 30 times, the time percentage of
failing to transmit data from any monitor to the
MFM-CMS does not exceed 0.1%.
Distinct Vision Distance The distinct vision distance between the monitor
and the AP: 50 meters.