Patient Monitor User Manual Default Settings
- 292 -
C Default Settings
This appendix documents the most important default settings of your monitor as it is delivered
from the factory.
NOTE:
If your monitor has been preconfigured according to your requirements, the settings at
delivery will be different from the default settings listed here.
C.1 Patient Information Default Settings
Patient Information Settings
Patient Type Adult
Pace Off
C.2 Alarm Default Settings
Alarm Settings
Pause Time 120 s
Sensor Off Alarm Off
Alarm Latch Off
C.3 ECG Default Settings
ECG Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 120 160 200
Alarm Low Limit 50 75 100
ARR Analysis ThresholdValue
Low Voltage(Limb) 0.5 mV
Pause 3 s
Sustain VT 30 s
PAC Bigeminy 8/min
Pauses/min High 8/min
PVCs High 10/min