Patient Monitor User Manual Monitoring ECG
- 105 -
the first valid and complete ST waveform appears.
In the ST View interface, the user can display the current waveform, baseline waveform, or the
both by selecting Real, Baseline or Real+Base. The user can also hide or display ST points by
selecting Hide Points or Show Points. Besides, the user can record and print the ST view.
In the ST View interface, the user can save ST segment through clicking Save ST SEG. Up to 20
groups of ST segments can be saved. When the 21
st
ST segment is saved, the earliest ST segment
will be deleted.
NOTE:
The ST baseline and ST segment will be cleared in following situations:
1) Turning off the monitor;
2) Changing the electrode type;
3) Changing the calculation lead in 3 electrodes;
4) Entering or exiting Demo mode;
5) Changing the patient type;
6) Admitting new patients;
In order to view the ST value situation of each lead more intuitively, the user can enter into the
ST Histogram. The horizontal axis shows the lead name while the vertical axis shows the ST
value. And the bar graph is used to display the ST value result. The ST histogram refreshes with
ST View synchronously.
9.9 Arrhythmia Monitoring
9.9.1 Arrhythmia Analysis
The arrhythmia algorithm is used to monitor ECG of adult patients in clinics, and detect the
changes of heart rate and ventricular rhythm, and also save arrhythmia events and generate alarm
information. Arrhythmia algorithm can monitor paced and non-paced patients. Qualified
personnel can use arrhythmia analysis to evaluate patient’s condition (such as heart rate, PVCs
frequency, rhythm and ectopic beat) and decide the treatment. Besides detecting change of ECG,
arrhythmia algorithm can also monitor patients and give proper alarm for arrhythmia.
The monitor can support 2 configurations for ARR analysis, basic ARR or advanced ARR (also
called Basic or Advanced). The default is advanced ARR. For configuration selection, please
contact service personnel of the manufacturer.
NOTE:
1 Advanced ARR is not available in USA.
2 The measured PVCs and Pause/min will be displayed in main interface. Pause/min
is only applicable to advanced ARR.
3 Advanced ARR is intended to be used with MFM-CMS 2.65 or above version.