Patient Monitor User Manual Product Specification
- 270 -
Range N
2
O 0~100 vol%
Halothane/ Enflurane/
Isoflurane
0~8 vol%
Sevoflurane 0~10 vol%
Desflurane 0~22 vol%
O
2
1
0~100%
Resolution
N
2
O, O
2
1
1%
CO
2
, AG 0.1%
AwRR
Measurement range
2 ~ 150 rpm
Measuring accuracy
±
1 bpm (120 bpm and below),
Not specified (120 bpm above)
Resolution
1 rpm
Accuracy
CO
2
± (0.2vol% + 2% of
reading)
Typical conditions:
Ambient temperature: (25
± 3) °C
Barometric pressure: (760
± 10) mmHg
Balance gas: N
2
Sample gas flowrate: 100
ml/min
N
2
O
± (2vol% + 2% of
reading)
Hal, Enf, Iso, Sev,
Des
±(0.15vol% + 5% of
reading)
O
2
1
± (1vol% + 2% of
reading)
CO
2
± (0.3vol% + 4% of
reading)
Claimed working
environment
N
2
O
± (2vol% + 5% of
reading)
Hal, Enf, Iso, Sev,
Des
±(0.2vol% + 10% of
reading)
O
2
1
± (2vol% + 2% of
reading)
Anesthetic
gas
identification
method
Manually set anesthetic gas
1
Automatic identification of an anesthetic gas
2
Automatic identification of two anesthetic gases
3
Anesthetic 1
st
gas identification concentration
2
> 0.2vol%