Patient Monitor User Manual Product Specification
- 274 -
Rise time (@ 50
sml/min sample flow)
CO
2
≤ 300 ms
N
2
O, O
2
, ENF, ISO , SEV, DES ≤ 400 ms
HAL ≤ 500 ms
(Measured according to EN ISO 80601-2-55)
Primary Agent
Threshold
0.15 vol%. When an agent is identified, concentrations will be
reported even below 0.15 vol% as long as apnea is not detected.
Secondary Agent
Threshold
0.2 vol% + 10% of total agent concentration
Agent Identification
Time
< 20 seconds (typically < 10 seconds)
Total system response
time
< 4 seconds (with 2 m Nomoline Airway Adapter Set sampling line)
Work Mode Measure
Data Update Period 1 s
Measurement Range CO
2
: 0 to 25 vol%
O
2
: 0 to 100 vol%
N
2
O: 0 to 82vol%
HAL, ENF, ISO, SEV, DES: 0 to 25 vol%
AwRR: 0 rpm to 150 rpm
Resolution CO
2
: 0.1%
HAL, ENF, ISO, SEV, DES: 0.1%
N
2
O: 1%
O
2
:1%
AwRR: 1 rpm
☆
Accuracy-
Standard
Conditions
Gas Range Accuracy
CO
2
0 to 15 vol%
15 to 25 vol%
±(0.2 vol% + 2% of reading)
Unspecified
N
2
O 0 to 82 vol% ±(2 vol% + 2% of reading)
HAL, ENF, ISO 0 to 8 vol %
8 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
SEV 0 to 10 vol %
10 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
DES 0 to 22 vol %
22 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified